TotaLab Automated Sample Preparation and Introduction Systems

Modular ICP-MS and ICP-OES sample preparation, calibration and introduction.

TotaLab5000-SD automated sample preparation and introduction system for semiconductor analysis

Automation & Robotics · 6 models

An ICP-MS or ICP-OES is only as good as the solution reaching its nebulizer. Calibration standards at parts-per-trillion levels are difficult to prepare by hand, high-matrix samples need a dilution an operator cannot reproduce exactly, and every transfer through a volumetric flask is a chance to add the contamination the instrument is meant to measure. TotaLab moves those operations into a valve-free, ultrapure PFA flow path between the autosampler and the plasma.

Valve-free, below 1 ppt

No valves in the wetted path, and ultrapure PFA throughout, hold the flow-path background under 1 ppt.

400-fold online dilution

In-line mixing dilutes high-matrix samples by more than 400 times, with a syringe pump stated to ±0.2%.

Prepare on demand

Standards are mixed online, so no volumetric flasks are cleaned and no excess high-purity reagent is consumed.

Cascade the modules

Each module runs standalone on an ICP-MS, or several can be chained into one preparation and introduction path.

Models

Choose the configuration.

The platform is modular by design. A single module can connect to an ICP-MS on its own, or several can be cascaded to cover calibration-curve preparation, online dilution, online preconcentration, online quality control and sample introduction in one chain. The same architecture serves semiconductor work, where electronic-grade wet chemicals, specialty gases and wafer surfaces are checked for trace metals and the flow-path background has to stay below 1 ppt.

01

TotaLab3000-SD3

Automated sample preparation and introduction system for ICP-OES and ICP-MS

TotaLab3000-SD3 combines sample introduction with online calibration, dilution and preconcentration for ICP-OES and ICP-MS. It is the general-purpose configuration in the line, aimed at laboratories running long element suites across metallurgical, chemical, food, agricultural and environmental samples. Sample trays hold 89 or 225 positions, and the needle is washed continuously in an overflow well between samples.

  • Automated sample introduction for ICP-OES and ICP-MS
  • Calibration curves prepared online from stock standards
  • Online sample dilution before introduction
  • Online preconcentration for low-level elements
  • 89 or 225-position sample trays with a dust cover
  • Dual continuous overflow needle wash, with two cleaning reagents
  • Wash solution, waste and sample-line liquid monitored during the run
Sample trays89 or 225 positions
Needle washDual continuous overflow, two reagents
02

TotaLab5000-SD

Sample preparation and introduction system for semiconductor trace-metal analysis

TotaLab5000-SD applies the platform to semiconductor materials: electronic-grade wet chemicals, electronic-grade specialty gases and wafer surfaces, checking them for metal impurities online. Its defining feature is a valve-free flow path built from PFA++ material, which holds the flow-path background below 1 ppt. It handles sample preparation and introduction as one sequence.

  • Built for electronic-grade wet chemicals, specialty gases and wafer surfaces
  • Valve-free flow-path design with background below 1 ppt
  • PFA++ wetted materials for trace-metal work
  • Online calibration and online quality control on the same path
  • Online specialty-gas absorption before introduction
  • 89, 96 or 225-position sample trays
  • Cascades with other TotaLab modules into a longer preparation chain
Flow-path backgroundBelow 1 ppt
Sample trays89, 96 or 225 positions
03

TotaLab5000-D

Automated online calibration-curve preparation system

TotaLab5000-D prepares calibration curves online, mixing standards from stock solutions as the run needs them. The purpose is to remove manual standard preparation from parts-per-trillion work, where a small volumetric error or a contaminated flask shifts the whole curve. Volumes are mixed in the PFA flow path, so no separate preparation vessels are used and only the required amount of standard is consumed.

  • Calibration curves mixed online from stock standards
  • Prepares only the volume needed, with no preparation vessels
  • Ultrapure PFA flow path with background below 1 ppt
  • Removes manual standard preparation from ppt-level work
  • Online quality-control samples added to the same sequence
  • Connects to an ICP-MS alone or as part of a cascade
Flow-path backgroundBelow 1 ppt
Curve formatOnline, mixed in the flow path
04

TotaLab5000-S

Automated sample introduction system

TotaLab5000-S is the introduction module in the 5000 series, feeding an ICP-MS or ICP-OES from high-capacity trays under software control. It carries the platform's sample protection features into a standalone autosampler: a dust cover over the rack, a continuous overflow needle wash, and monitoring of wash solution, waste and the sample line, so a fault surfaces during the run rather than in the results.

  • Automated sample introduction for ICP-MS and ICP-OES
  • 89, 96 or 225-position sample trays
  • Dual continuous overflow needle wash, switching two cleaning reagents
  • Dust cover over the sample rack
  • Wash solution and waste level monitoring
  • Sample-line liquid monitoring
Sample trays89, 96 or 225 positions
Needle washDual channel, two reagents
05

TotaLab3000-S

Automated sample introduction system

TotaLab3000-S is the introduction module in the 3000 series, for laboratories that want automated feeding and washing without the preparation functions of the SD models. It takes 89 or 225-position trays, washes the needle continuously in an overflow well with either of two cleaning reagents, and monitors the wash solution, the waste container and the liquid in the sample line.

  • Automated sample introduction for ICP-OES and ICP-MS
  • 89 or 225-position sample trays with a dust cover
  • Dual continuous overflow needle wash, two reagents
  • Wash solution and waste level monitoring
  • Sample-line liquid monitoring
  • Can be combined with the D-series preparation modules
Sample trays89 or 225 positions
06

TotaLab3000-D1/D2/D3

Online calibration, dilution and preconcentration modules

The D1, D2 and D3 variants add preparation functions to the 3000 series instead of a tray and needle, covering online calibration-curve preparation, online sample dilution and online sample enrichment. They use the same in-line mixing approach and PFA flow path as the rest of the platform, and can be connected to an ICP-MS individually or cascaded with the introduction modules.

  • D1, D2 and D3 configurations for online curve preparation, dilution and preconcentration
  • In-line mixing for dilution of high-matrix samples
  • Dilution capability beyond 400-fold
  • Syringe pump accuracy of ±0.2%
  • Ultrapure PFA flow path with background below 1 ppt
  • Runs alone with an ICP-MS or cascaded with other modules
Dilution capabilityMore than 400-fold
Syringe pump accuracy±0.2%
Flow-path backgroundBelow 1 ppt

Full specifications for other configurations, voltages and options are available on request.

The range

Configurations and accessories.

Applications

Where this range is used.

  • Trace metals in electronic-grade wet chemicals and specialty gases
  • Wafer-surface metal contamination checks
  • Calibration curves for ICP-MS at parts-per-trillion levels
  • Online dilution of high-matrix digests and metallurgical solutions
  • Online preconcentration of trace elements in water samples
  • High-throughput element suites in food, soil and agricultural digests

Semiconductor & High-Purity Materials →

Questions

Before you buy.

What does a valve-free flow path change in trace-metal analysis?

Valves add dead volume, wetted surfaces and places for contamination to collect. Removing them from the wetted path, together with PFA or PFA++ tubing, is how the platform reaches a stated background below 1 ppt, which matters when the analytes themselves sit at that level.

Can TotaLab prepare calibration curves online instead of by hand?

Yes. The D models and the SD systems mix standards from stock solutions in the flow path, so no volumetric flasks are filled or cleaned. Because only the required volume is prepared, high-purity reagent consumption falls, and the operator is out of the ppt-level dilution chain.

Which instruments do TotaLab modules connect to?

The 3000 and 5000 series are specified for ICP-OES and ICP-MS. A single module can connect to an ICP-MS on its own, or several modules can be cascaded, and the semiconductor configuration adds electronic-grade wet chemicals, specialty gases and wafer-surface work.

Send the matrix and the limit

We reply in one working day with a configuration, the consumables it needs and a lead time.